Listed Below
 
Request Scruz News
Thickness Gauges
Flaw Detectors
Bolting Products
Transducers
Transducers
Transducers
Transducers
Transducers
Return to Main Page
Group Logo
 
 
 
 

The MMX-6 is a simple to use hand-held Ultrasonic Thickness Gauge with the ability to measure through paint and coatings and eliminate the thickness of the paint or coating. The MMX-6 uses a dual element style transducer. With the single press of a button, the MMX-6 can be switched between pit or flaw mode ( pulse-echo), and through paint/coatings mode (echo-echo) for maximum inspection efficiency. The MMX-6 is equipped with a bundle of features to make your job eaiser: Alarm mode, high speed scanning, and data send are the main features of the MMX-6.  The MMX-6 comes as a complete kit, ready to use, and is backed by Dakota Ultrasonics 5 year limited warranty.

 
Applications:
 

Corrosion & Pitting
Tube & Pipe
Tanks
Boilers
Glass
Variety of Applications
 
Specifications:
 
  

Physical 

Weight:  10 ounces 

Size: 
2.5 W x 4.75 H x 1.25 D inches (63.5 W 120.7 H x   31.8 D mm) 

Operating Temperature: 
-20° to 120° F (-30° to 50° C) 

Case:
Extruded aluminum body / nickel plated aluminum end caps.

Resistant to impact and environmentally sealed.

Keypad 
Sealed membrane that is  resistant to both water and  petroleum products.
 
Nine tactile-feedback keys. 

Transducer
 

Dual element (transmit and  receive). 

1 to 10 MHz frequency range.

Special high damped transducers available for use with through paint or coating measurements).

Locking quick disconnect LEMO connectors.
 
4 foot cable.
 
Optional cable lengths up to 100 feet. 

Custom transducers available for special 
applications. 

  

  

  

 

  

Power Source 

Two 1.5V alkaline or  1.2V NiCad AA cells 

Typically operates for  200 hours on alkaline and 
130 hours on NiCad. 

Display flashes when battery is low. Unit turns off automatically when battery  is too low to operate reliably. 

Display 

Multi-function 4.5 digit liquid  crystal display with 0.500 inch numerals, backlit for use in poor light conditions.  

Backlight is selectable:  
on / off / auto ( illuminates only when taking a  
measurement ). 

Measurements displayed in  inches, inches/microsecond.  millimeters, and mters/second. 

Bar graph indicates stability 
of reading. 

Data Logger

Automatic numeric Data Logging system 1 to 1000 readings.

Certification 
Factory calibration traceable  to national standards. 

Warranty 

5 year limited

  

Measuring

Range:

Pulse Echo Mode(Pit & Flaw Detection):
Measures from 0.025 to 19.999 inches (0.63 to 
500 millimeters).

Echo Echo Mode (Thru Paint & Coatings):
Measures from 0.100 to 1.0 inches (2.54 to 25.4 millimeters).

Switch  to select English or metric units.
 

Resolution: 
0.001 inches (0.01 millimeters) 

Velocity Range:
.0492 to .3937 in/m s
1250 to 10,000 meters/sec

Built in: 

Stainless steel reference disk for probe zeroing. 

Four readings per second  for single point measurements or 16 per second in scan mode – captures the minimum thickness. 

Single and Two point calibration option included. 

Features

Multi-Mode: Toggle between Pit/Flaw detection (Pulse Echo Mode), and through Paint/Coatings (Echo-Echo Mode) with the single press of a button.
 

Alarm Mode:
Enter a minimum acceptable thickness value. 
  If measurement falls  below minimum, red led will illuminate and sound beeper.  If measurement  is above minimum value, green led will illuminate. 

Data Ouput:
RS232 output sends thickness
  readings to a printer, computer or other external storage device.  

 

 

 



   
 
Larger Image
Click on the photo above for a larger image.
 
 
 
 
Powered By:
Dakota Ultrasonics 2008
Live chat by Boldchat
 
ASNT Society
 
NACE Society
 
NDT Solutions
 
SonaTest
 